Specifications

 

Spectrometer  
Type 4-channel polychromator + reference channel
Optical principle Ultrasensitive photodiode array spectrometer for UV-VIS range
Optical arrangement Catadioptric Czerny-Turner (Patent US 7,864,317 B2)
Light source 6W Xenon flash lamp lifetime 109 flashes
Wavelength range 220-750 nm
Wavelength resolution < 3 nm
Wavelength accuracy 0.5 nm
Wavelength reproducibility       0.1 nm
Photometric range 0.0005 – 2.0 OD
Absorbance precision 0.003 OD
Full spectrum acquisition 20 per second
AD conversion 16 bit
   
Plates Compatibility       DropPlates 16S, 16D : up to 6 using the aluminium frame
  DropPlates 96S, 96D (SBS format, no holder required)
 

Commercial 96-well UV-Vis compatible plates

(contact Trinean technical support to check feasibility for your application)

   
Other Technical Data  
System start-up time < 2 minutes including full system self-test
Operation Computer controlled (Windows)
Software compatibility DropQuant and cDrop (all versions)
Communication USB
Power supply 110 – 240 V, 50/60Hz (Operating voltage 24 VDC)
Size 34 x 45 x 35 cm (WxDxH)
Weight 30 kg
Compliance CE (and CSA in near future)